Reproducibility of the interpretation of pelvic x-ray 3 months after hysteroscopic sterilization with Essure

  • Sebastiaan Veersema, M.D.

      Affiliations

    • St. Antonius Hospital, Nieuwegein, the Netherlands
    • Corresponding Author InformationReprint requests: Sebastiaan Veersema, M.D., Koperwiekweg 34, 3403 ZS IJsselstein, the Netherlands (FAX: + 31-30-6874669).
  • ,
  • Ben W.J. Mol, M.D., Ph.D.

      Affiliations

    • Academic Medical Centre, Amsterdam, the Netherlands
  • ,
  • Hans A.M. Brölmann, M.D., Ph.D.

      Affiliations

    • VU University Medical Center, Amsterdam, the Netherlands

Received 23 September 2008; received in revised form 10 May 2009; accepted 11 May 2009. published online 22 June 2009.

Objective

To estimate the diagnostic accuracy and the interobserver reproducibility of pelvic x-rays in the diagnosis of successful bilateral sterilization with Essure after a 3-month follow-up period.

Design

Interobserver study.

Setting

Outpatient department of obstetrics and gynecology in a Dutch teaching hospital.

Patient(s)

Patients with successful bilateral Essure placement.

Intervention(s)

Hysteroscopic sterilization with Essure and pelvic x-ray and hysterosalpingography after a 3-month follow-up period.

Main Outcome Measure(s)

Six observers evaluations of 47 pelvic x-rays from 47 patients 3 months after a technical successful bilateral placement of microinserts to estimate the reliability of the sterilization. Diagnostic accuracy of pelvic x-ray per observer in detecting incorrectly positioned microinserts was expressed in terms of sensitivity and specificity, with hysterosalpingography as the reference strategy. Reproducibility of the interpretation of the pelvic x-ray was expressed as κ-values.

Result(s)

The sensitivity and specificity for x-rays read by gynecologists was 0.67 (95% confidence interval [CI], 0.29–0.96) and 0.79 (95% CI, 0.58–1.00) and for radiologists 1.0 and 0.5 (95% CI, 0.36–0.64). The interobserver agreement in reliability of pelvic x-ray of hysteroscopic sterilization assessment with Essure ranged from slight (κ-value = 0.09) for gynecologists to moderate (κ-value = 0.52) for radiologists.

Conclusion(s)

Test characteristics of pelvic x-ray as the imaging technique to assess the position of the Essure microinserts and tubal patency were poor, as was the reproducibility, particularly if gynecologists performed the evaluation. We do not recommend the use of pelvic x-ray for the assessment of the positioning of microinserts after hysteroscopic sterilization.

Key Words: Essure, hysteroscopic sterilization, confirmation test, x-ray, interobserver reproducibilty

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 S.V. is a consultant for Conceptus SAS and a member of the advisory board for Hologic Inc. B.W.J.M. has nothing to disclose. H.A.M.B. has nothing to disclose.

PII: S0015-0282(09)01099-1

doi:10.1016/j.fertnstert.2009.05.023

Fertility and Sterility
Volume 94, Issue 4 , Pages 1202-1207, September 2010